It’s a question I’m asked probably once a week. “What wavelength do I select on my luminometer when performing a luciferase assay?” The question is a good and not altogether unexpected one, especially for those new to bioluminescent assays. The answer is that in most cases, you don’t and in fact shouldn’t select a wavelength (the exception to this rule is if you’re measuring light emitted in two simultaneous luciferase reactions). To understand why requires a bit of an explanation of absorbance, fluorescence, and luminescence assays, and the differences among them.
Absorbance, fluorescence, and luminescence assays are all means to quantify something of interest, be that a genetic reporter, cell viability, cytotoxicity, apoptosis, or other markers. In principle, they are all similar. For example, a genetic reporter assay is an indicator of gene expression. The promoter of a gene of interest can be cloned upstream of a reporter such as β-galactosidase, GFP, or firefly luciferase. The amount of each of these reporters that is transcribed into mRNA and translated into protein by the cell is indicative of the endogenous expression of the gene of interest.
Wavelength for Absorbance Assays
β-galactosidase assays are traditionally measured by absorbance. The principle of the β-galactosidase assay is that a specific substrate is converted into a specific compound by the β-galactosidase enzyme. The product of that reaction absorbs light maximally at a specific wavelength. But the principle is the same for other compounds of interest for other absorbance assays. The more light that is absorbed, the more of that compound is present in a sample. But because biological samples are complex mixtures of many different compounds that absorb light at different wavelengths, you need to filter out the other wavelengths that might also be absorbed by compounds you don’t want to measure, and would otherwise give an inaccurate measurement.
Wavelength for Fluorescence Assays
A fluorescence assay that uses GFP as a genetic reporter is similar. More GFP means greater expression of the gene of interest. To measure GFP or other fluorescent compounds, a specific wavelength of light is directed at a sample. Light reemitted by GFP or other fluorescent compounds is measured by a fluorometer. For fluorescent compounds, there is both a maximal excitation wavelength (the wavelength of light directed at a compound that will result in the highest amount of light reemitted at a different wavelength) and a maximum emission wavelength (the wavelength of light emitted by the fluorescence compound with the highest intensity). So measuring fluorescence requires two wavelengths be selected. The first filters out light that might excite other fluorescent compounds present in a sample, and the second filters out all light except the wavelength of light emitted by GFP or a different fluorescent compound of interest. Fluorescence assays are more sensitive than colorimetric assays. But because so much of the light must be filtered for an accurate measurement, their signal-to-noise or background is high.
Why You Usually Don’t Need to Select a Wavelength for a Luminescent Assay
Luminescence assays, on the other hand, are different because no light is directed at the sample. In a firefly, Renilla, or NanoLuc® luciferase assay, light is produced as a by-product of a reaction in which a luciferase enzyme converts a substrate to a product. The amount of light produced is proportional to the amount of the luciferase enzyme and the level of expression of the gene of interest if used as a genetic reporter. Most eukaryotes don’t produce light, so in general, the background light from a sample is very low. This means that there is no need to filter out any light coming from a luciferase assay. In fact, doing so means some of the sensitivity of the assay will be lost because while luciferase reactions will emit light maximally at specific wavelengths, the light emitted from a luciferase reaction is quite broad across many wavelengths. In fact, most luminometers will collect light from the entire visible spectrum. This results in a very sensitive assay with low background.
So the next time you need to perform a luciferase assay, remember, in most cases you don’t need to select a specific wavelength.
Additional reading is proved in the following resource:
Designing a Bioluminescent Reporter Assay: Normalization
To explore a world of possibilities for bioluminescent assays, click here.
Related Posts
Updated 2/21/21
Ben Schmidt
Latest posts by Ben Schmidt (see all)
- Tips for Successful Dual-Reporter Assays - December 6, 2019
- Why You Don’t Need to Select a Wavelength for a Luciferase Assay - June 7, 2019
- To inject or not inject? - September 15, 2014
Thank-you for writing this. It now seems very obvious, but I was getting frustrated trying to figure out why I couldn’t find a wavelength for a bioluminscence assay that I wanted to do. This was very helpful!
Glad that you found this blog useful.
Is this just for firefly, Renilla, or NanoLuc® luciferase assay or this is the rule for any luciferase assay. I need to measure ATP on luminometer with luciferase-luciferin assay. We have luminometer Fluoroscan Ascent FL, but when I change to luminometer mode, I have menu for filter… and there is writen NONE.
In assay for ATP measurement is writen that luminiscence is monitored at 560nm.
For luminescence you should not need a specific wavelength, because unless you are using measuring different luciferase reporters at the same time. I don’t know why your specific assay is suggesting a wavelength, but it might be worth a chat with our technical services scientists about your particular application. They can be contacted by email or online chat: https://www.promega.com/support/customer-and-technical-support/
Thank you, this short review was very useful for me. :)
Does EGFP interfere the luminescence assay, given that the spectrum of luminescence is broad (may emit blue light?)
Even though there is no external excitation of light, light produced from luminescence can be absorbed by fluorescent protein inside cells (if any), and can mask the actual luminescence ? Theoretically, it is a possibility right?
Does a Promega GloMax Jr. detect infra red produced by firefly luciferase?
This explanation is so clear that correct my “unclear” understanding several years back
Is there a colorimetric substrate available for lucferase? Not luminescence.
DO I have to set filters if I’m going to use the DUAL GLO LUCIFERASE ASSAY?
The Dual-Glo Luciferase Assay is measuring luminescent assay, so you will not need to “excite” the sample to get the glow reaction. Firefly and Renilla luciferases require different substrates for their reactions. The protocol allows you to add the firefly luciferase substrate and measure signal firefly then add the Stop&Glo reagent which quenches the firefly luciferase signal and provides reagents required for the Renilla signal. IT IS VERY IMPORTANT to normalize your results to controls. The process is described in the Technical Manual TM058 https://www.promega.com/-/media/files/resources/protocols/technical-manuals/0/dual-glo-luciferase-assay-system-protocol.pdf?rev=e76815ff4e3a4aa29b969d62a2b11191&sc_lang=en and the appendix of the manual contains a nice explanation of how the signals differentiatied and how to normalize to control wells. If you have any additional questions, you can contact one of our technical services scientists at techserv@promega.com or at https://www.promega.com/support/